Parameters of the Axesse® linacs

Parameter Axesse®
Leaf width at isocenter (mm) 5
Field size, maximum (mm) 400
Leaf individual travel range (mm) 200 (with respect to DLGa)
Leaf and DLGa or backup jaw combined travel range (mm) 350
Leaf interdigitation range (mm) 200
Diaphragm over-travel relative to central axis (mm) 120
Height of leaf (mm) 90
Leaf and interleaf transmissions (6MV) 0.47 %
Diaphragm/jaw speed, maximum (mm/s) 90
Leaf speed, maximum (mm/s) 35
Leaf and DLG combined speed, maximum 65 mm/s
Speed of gantry rotation (/s) 6.0 degree/s
Dose rate, maximum (MU/min) Continuous, 660

a DLG Dynamic Leaf Guides. All leaves were integrated with two DLGs and they traveled together