Table 3.

Mean CD4+ cell counts and uncertainties derived for different vials and dilutions

Non weighted mean value Inline graphic μL−1 Uncertainty of mean value Inline graphic μL−1 Standard deviation Inline graphic μL−1 Number of measurement values n Expansion factor k Inline graphic Consistency test
Dilution 1, vial 1 327.9 9.1 39.6 21 2.080 243.1 Negative
Dilution 1, vial 2 297.4 7.4 31.0 20 2.086 445.7 Negative
Dilution 1, vial 3 288.3 10.0 43.5 20 2.086 495.4 Negative
Dilution 2, vial 1 307.7 6.3 26.8 22 2.074 158.6 Negative
Dilution 2, vial 2 305.3 6.0 25.0 22 2.074 105.1 Negative
Dilution 2, vial 3 284.4 9.9 44.8 22 2.074 374.3 Negative

Values for each of six vial / dilution combinations were averaged over all participants.