. 2015 Nov 16;42(12):7034–7042. doi: 10.1118/1.4935530

TABLE II.

The CT numbers measured from the physically acquired ACR phantom image, “commercial,” and the CT numbers measured from the simulated ACR phantom image, “forward projection.” Unit: HU. The air, acrylic, bone, polyethylene, and solid water materials are provided by inserts in Module 1 of ACR phantom.

Air Acrylic Bone Polyethylene Solid water
80 kV Commercial −987 102 1178 −111 3
Forward projection −988 103 1180 −112 4
100 kV Commercial −987 115 980 −93 2
Forward projection −988 116 981 −94 2
120 kV Commercial −987 121 868 −83 3
Forward projection −987 121 868 −82 4
140 kV Commercial −985 125 797 −75 3
Forward projection −985 126 796 −76 4