Skip to main content
. 2014 Mar 1;12(2):110–119. doi: 10.1089/adt.2013.552

Table 1.

Acceptance Criteria for Wells

Parameter Acceptance criterion
RSEAL (baseline)a >300 MΩ
Current amplitude (baseline) >0.2 nA
RSEAL stability <50% decrease
Current stability <50% decrease in mean of vehicle control
Voltage clamp quality Visual inspection
a

Typical RSEAL values ranged between 500 and 1,000 MΩ.