Skip to main content
. 2014 Mar 1;12(2):110–119. doi: 10.1089/adt.2013.552

Table 2.

Acceptance Criteria for 384-Well Population Patch Clamp Plate

Parameter Acceptance criterion
Z′ factor ≥0.4
SW ≥2
Success rate >80% accepted wells per PPC plate
IC50 for reference inhibitors ≤0.5 log from historical mean

SW, signal window; PPC, population patch clamp.