. 2015 Nov 26;10:175. doi: 10.1186/s13019-015-0376-9

Table 2.

Postoperative and follow-up results after successful device closure of doubly committed juxtaarterial ventricular septal defects

IADR (n = 58) At discharge (n = 58) 1 Months (n = 58) 6 Months (n = 58) 12 Months (n = 42) 24 Months (n = 28) 36 Months (n = 14)
Pre-existing AR (n = 16) 12 9 7 6 4 2 1
New AR (n) 0 0 0 0 0 0 0
New mild PR (n) 2 2 2 2 1 1 0
Complete Closure (%) 50/58 (86) 55/58 (95) 57/58 (98) 57/58 (98) 41/42(98) 27/28(96) 14/14(100)
Trivial RS (%) 6/58 (11) 1/58 (2) 0 0 0 0 0
Small RS (%) 2/58 (3) 2/58 (3) 1/58 (2) 1/58 (2) 1/42 (2) 1/28(4) 0
Follow-up median (days) 880 (180–1860)

IADR immediately after device release; AR aortic regurgitation; PR pulmonary regurgitation; RS residual shunts