Table 3.

Optimum collimator configuration & BiCoreTM collimators specifications.

Isotope Proposed Energy-optimized Collimator
LEHR
99mTc
MELP
67Ga
57Co 99mTc 123I 111In
Hole shape Square Square Square Square Hexagonal Hexagonal
Number of Holes 16384 16384 16384 16384 148000 14000
Hole length 23 mm 23 mm 23 mm 23 mm 24.05 mm 40.64 mm
Septal thickness 0.32 mm 0.32 mm 0.32 mm 0.32 mm 0.16 mm 1.14 mm
Hole material Tungsten alloy Tungsten alloy Tungsten alloy Tungsten alloy Lead Lead
Hole Diameter 1.28 mm 1.28 mm 1.28 mm 1.28 mm 1.11 mm 2.94 mm
Sensitivity @10 cm 291 cpm/uCi 261 cpm/uCi 241 cpm/uCi 504 cpm/uCi 202 cpm/uCi 310 cpm/uCi
Resolution @10 cm 6.82 mm 6.91 mm 7.08 mm 7.29 mm 7.4 mm 12.5 mm
Septal Penetration 3.7% 4.1% 11.3% 17.9% 1.5% 1.2%