Table 3.
Isotope | Proposed Energy-optimized Collimator |
LEHR 99mTc |
MELP 67Ga |
|||
---|---|---|---|---|---|---|
57Co | 99mTc | 123I | 111In | |||
Hole shape | Square | Square | Square | Square | Hexagonal | Hexagonal |
Number of Holes | 16384 | 16384 | 16384 | 16384 | 148000 | 14000 |
Hole length | 23 mm | 23 mm | 23 mm | 23 mm | 24.05 mm | 40.64 mm |
Septal thickness | 0.32 mm | 0.32 mm | 0.32 mm | 0.32 mm | 0.16 mm | 1.14 mm |
Hole material | Tungsten alloy | Tungsten alloy | Tungsten alloy | Tungsten alloy | Lead | Lead |
Hole Diameter | 1.28 mm | 1.28 mm | 1.28 mm | 1.28 mm | 1.11 mm | 2.94 mm |
Sensitivity @10 cm | 291 cpm/uCi | 261 cpm/uCi | 241 cpm/uCi | 504 cpm/uCi | 202 cpm/uCi | 310 cpm/uCi |
Resolution @10 cm | 6.82 mm | 6.91 mm | 7.08 mm | 7.29 mm | 7.4 mm | 12.5 mm |
Septal Penetration | 3.7% | 4.1% | 11.3% | 17.9% | 1.5% | 1.2% |