Table 1. Comparison of T2 Estimation Accuracy for REPCOM and Echo Sharing Based on a Physical Phantom Composed of Cylindrical Vials Embedded in a Background Material with T2 = 45 ms.
Object | Reference | REPCOM |
Echo Sharing |
||
---|---|---|---|---|---|
Diameter (mm) | T2 (ms) | % Error | % Error | ||
5 | 233.2 | 233.7 | 0.2 | 188.1 | −19.3 |
5 | 167.3 | 166.9 | −0.2 | 160.0 | −4.4 |
5 | 83.2 | 83.8 | 0.7 | 89.9 | 8.0 |
10 | 231.9 | 235.6 | 1.6 | 212.4 | −8.4 |
10 | 168.2 | 168.6 | 0.2 | 159.0 | −5.5 |
10 | 83.8 | 83.4 | −0.5 | 85.5 | 2.0 |