Table 6. Summary of outcomes of the three studies evaluating individualized preformed implants.
| Study | % of patients experiencing residual diplopia | Enophthalmos | Complications | 
|---|---|---|---|
| Nkenke22 | N/A | Reduction in Enophthalmos (mm) ± SD Assisted: 3.25 ± 1.44 Control: 3.88 ± 1.23 (p = 0.31) | Assisted: 1 reduced vision, 2 suspected retrobulbar hematoma Control: 2 suspected retrobulbar hematoma (p = N.S.) | 
| Scolozzi23 | N/A | Reduction in Enophthalmos (mm) ± SD Assisted: 1.5 ± 1.58 Control: 1.8 ± 1.61 (p = 0.79) | N/A | 
| Guo24 | Assisted: 17% Control: 88% (p = 0.0042) | Assisted: 21%* Control: 50%* (p = 0.04)* | N/A | 
N/A = not available; N.S. = non-significant; *percentage of patients having residual enophthalmos; *exophthalmometry data not reported; SD = standard deviation.