Published in final edited form as: J Clin Densitom. 2015 Jun 10;19(2):220–225. doi: 10.1016/j.jocd.2015.04.003

Table 1.

Phantom and Human In Vivo Data and Comparisons on the Existing and New Instrument

GE Spine Phantom L1-L4 BonaFide Phantom (BFP) L1-L4 Hologic Phantom L1-L4 Bioclinica Body Composition Phantom (BBCP) Total Body Human In Vivo Total Body
Bias Existing Unit Mean New Unit Mean Bias Existing Unit Mean New Unit Mean Bias Existing Unit Mean New Unit Mean Bias Existing Unit Mean New Unit Mean Bias Existing Unit Mean New Unit Mean
BMD g/cm2 −0.007 1.188 1.195 −0.008 1.070 1.079 −0.003 1.105 1.108 0.005 1.081 1.086 0.026 1.250 1.276
Area cm2 0.252 51.29 51.04 0.269 44.77 44.51 0.512 52.46 51.95 0.067 497.6 497.7 −11.3 2432 2420
BMC g −0.062 60.93 61.00 −0.085 47.92 48.00 0.425 57.99 57.56 −3.367 537.7 541.1 −50.9 3051 3102
Lean g - - - - - - - - - 462.9 10534 10071 832.9 54230 53397
Fat g - - - - - - - - - −496.4 5924 6420 −860.2 17507 18267
Fat % - - - - - - - - - −2.84 34.9 37.7 −1.14 23.4 24.6