Table 2.

Total emboli volume versus circuit type, flow rate and pressure, and transducer.

Volume (mL)
Circuit Pressure/Flow N Transducer 1 Transducer 2 Transducer 3
SX/D734 A  15 4.2 (1.2) .025 (.022) .00078 (.002)
B 15 8.8 (2.2) .084 (.082)   .0013 (.0017)
RX/AF125 A 15 4.9 (1.3) .072 (.036)   .0015 (.0012)
B 15  11 (3.8)   .25 (.079)    .013 (.0075)
FX A 15 2.6 (.81) .046 (.036)  .0040 (.0037)
B 15 4.5 (2.2) .16 (.11)  .039 (.054)
SYN102 A 15  15 (8.7) .58 (.25)  .38 (.25)
B 15  17 (9.9) 1.1 (.34)  .69 (.27)
SYN40 A 15   .16 (.048) .000081 (.00016)   .00032 (.0005)  
B 15 1.0 (.51) .0091 (.026)   .0057 (.018)  

Raw data presented as mean × 10−5 (±standard deviation × 10−5).

A: Flow rate of 3.5 L/min and a pressure of 150 mmHg.

B: Flow rate of 5.0 L/min and a pressure of 200 mmHg.