Table 1.

Relative error of CDI estimates with and without gas calibration.

With Gas Calibration Without Gas Calibration
Before ABL calibration pH .1 ± .2% 9.0 ± 2.6%*
pCO2 −2.9 ± 5.2% −26.5 ± 8.8%*
pO2 .3 ± 4.5% −37.8 ± 5.6%*
After ABL calibration pH .2 ± .2% .1 ± .2%
pCO2 −1.0 ± 2.0% −1.1 ± 2.2%
pO2 −.9 ± 7.8% −.5 ± 8.1%

Values are mean ± SD.

*

Mean values are significantly different from zero (p < .05).