Skip to main content
. 2009 Mar;41(1):20–27.

Table 2.

MB counts per minute (control) + 500 mL/min vent

Before the Membrane (MB/min) After the Arterial Filter (MB/min) Percent Decrease p
Vision 315 ± 118   3.0 ± 2.0 99    .0100
RX25 668 ± 31     7.0 ± 5.0 99    .0008
Synthesis 464 ± 192   4.0 ± 2.0 97.6 .0143
Affinity 670 ± 9.0  131 ± 97 80.4 .0106
Quadrox 685 ± 6.0  343 ± 66 49.9 .0124