Table 2.
In Vitro Test | Ranking Criteria | Comment |
---|---|---|
Separation of flowing GME* | Percent removal at rated pore size | The percent removal of flowing GME distribution* below the reported pore size (μm) of the filter screen |
90% efficiency rating | Micrometer size where filter removes 90% of flowing GME distribution* | |
Room air bolus | Largest micrometer gas emboli passed | Largest micrometer gaseous emboli measured at filter outlet after room air bolus |
Cubic micrometers of gas passed after room air bolus | Gas volume passed in ×107 μm3 after filter inlet 10 mL room air bolus calculated from outlet GME distribution | |
Seconds after bolus | Seconds that GME persist down to 2 GME/sec at filter outlet after room air bolus | |
Blood flow resistance | Prime volume in cubic centimeters | The prime volume reported by the manufacturer |
Flow resistance (mmHg/L/min) | Calculated from pressure drop measurements at 4.5 L/min for bovine 20% hematocrit blood at 32°C |
Flowing GME distribution is >104 10- to 200-μm suspended gaseous emboli flowing at 4.5 L/min, 20% hematocrit bovine blood at 32°C.