Skip to main content
. 2006 Sep;38(3):206–213.

Table A2.

Bubble oxygenator GME outlet counts at approximate micron size ranges for five different test solutions filtered and not filtered.

Solution 0–15 μm 16–30 μm 31–45 μm 46–60 μm 61–75 μm 76–90 μm
GME counts at approximate
  micron size up to 90 μm
      20% glycerin 131* 237* 106*   18*     1*   0
      28% glycerin 206   519   473   266   63 13
      23% hematocrit 235   607   469   191   39   5
      32% glycerin   53* 355* 407* 380* 274* 117*
      37% hematocrit   98* 547   527* 435* 204*   48*
GME counts at approximate
  micron size up to 38 μm
      20% glycerin 69 136*   48*     6*    1   0
      28% glycerin 150   467   399   99    9   1
      23% hematocrit 290   520   166   14    1   0
      32% glycerin 105   469   460* 322*    93*   20*
      37% hematocrit 167   581* 359* 19    3   0

N = 3–10 observations per cell.

*

Compared with 23% hematocrit P < 0.05.