Table 1. Characterization of MCO hemodialysis membranes, conventional high-flux and high cut-off membranes, based on dextran sieving experiments before and after blood exposure.

  Before blood exposure
After blood exposure
Membrane MWRO [kDa] MWCO [kDa] Pore radius [nm] MWRO [kDa] MWCO [kDa] Pore radius [nm]
Revaclear 5.7 ± 0.5 32 ± 3 3.9 ± 0.1 4.4 ± 0.3 14.2 ± 0.2 2.68 ± 0.02
MCO 1 9.4 ± 0.2 56 ± 3 5.0 ± 0.1 5.0 ± 0.4 18.1 ± 0.8 3.0 ± 0.1
MCO 2 10.0 ± 0.6 64 ± 3 5.4 ± 0.1 5.84 ± 0.09 18.2 ± 0.3 3.0 ± 0.1
MCO 3 11.3 ± 0.4 81 ± 9 6.0 ± 0.3 6.32 ± 0.06 22.7 ± 0.8 3.3 ± 0.1
MCO 4a 12.1 ± 0.7 99 ± 7 6.5 ± 0.2 6.77 ± 0.06 25 ± 5 3.5 ± 0.3
Theralite 15 ± 1 300 ± 100 10 ± 2 8.1 ± 0.8 40 ± 8 4.3 ± 0.4

Values are average ± standard deviation for n = 3.

aExperiments for MCO 4 included also one dextran fraction of 150 kDa.