Table 4. Fitted results from phantom measurements a MSE sequence with different FAs as used in the sequence protocol (see Supporting Information, S1 Dataset: MSE datasets) and different fitting methods.
Method 1 | Method 2 | ||||
FA | T2 (ms) | dT2 (%) | T2 (ms) | dT2 (%) | Offset (a.u) |
180° | 70.7 ± 1.0 | 3.5 ± 0.8 | 69.4 ± 1.2 | 1.9 ± 0.8 | 33949.6 |
160° | 72.2 ± 0.9 | 5.6 ± 1.1 | 70.9 ± 1.1 | 4.1 ± 0.9 | 28239.8 |
140° | 77.0 ± 0.7 | 12.7 ± 1.4 | 76.0 ± 0.9 | 11.6 ± 1.4 | 36127.6 |
120° | 84.7 ± 0.6 | 24.0 ± 1.9 | 84.5 ± 0.8 | 24.1 ± 1.9 | 37770.0 |
Method 3 | Method 4 | ||||
FA | T2 (ms) | dT2 (%) | T2 (ms) | dT2 (%) | Offset (a.u) |
180° | 70.6 ± 1.0 | 4.6 ± 1.0 | 69.3 ± 1.1 | 3.6 ± 0.7 | 27230.2 |
160° | 71.3 ± 0.9 | 5.5 ± 1.0 | 68.9 ± 1.2 | 2.8 ± 0.8 | 33127.8 |
140° | 73.7 ± 0.8 | 9.1 ± 1.0 | 71.1 ± 1.1 | 6.2 ± 0.8 | 30207.4 |
120° | 78.2 ± 0.8 | 15.7 ± 1.3 | 74.5 ± 1.1 | 11.3 ± 1.0 | 21556.9 |