Skip to main content
. Author manuscript; available in PMC: 2016 Nov 1.
Published in final edited form as: Physiol Meas. 2015 Oct 9;36(11):2301–2317. doi: 10.1088/0967-3334/36/11/2301

Table 5.

Doppler Wire vs. Flowmeter Linear Regression Statistics.

Straight Segments

r2 Slope
95% CI
Intercept (mL/s)
95% CI
Standard Error
of Regression (mL/s)

Time Average .957 1.07
[.93, 1.21]
−.50
[1.04, 0.03]
.216
Peak .962 1.15
[.92, 1.39]
.15
[1.61, 1.92]
.539

Aneurysm Models

r2 Slope
95% CI
Intercept (mL/s)
95% CI
Standard Error
of Regression (mL/s)

Time Average .900 1.05
[.82, 1.28]
−.76
[−1.63, 0.11]
.322
Peak .865 1.31
[.97, 1.66]
−2.36
[−4.95, 0.23]
.694