. 2015 Dec 22;2(1):e000225. doi: 10.1136/openhrt-2014-000225

Table 3.

Comparison of IVUS and FD-OCT assessments following rotational atherectomy (n=18)

IVUS FD-OCT Difference (IVUS—FD-OCT) p Value
Mean reference lumen diameter, mm 3.64±0.80 3.38±0.66 0.26±0.34 0.330
Mean reference lumen area, mm2 10.77±1.14 9.49±3.59 1.28±1.69 0.388
In-stent diameter, mm
 Minimum 2.51±0.53 2.48±0.48 0.02±0.17 0.586
 Maximum 4.38±0.82 3.81±0.67 0.57±0.28 <0.001
 Mean 3.38±0.62 3.25±0.60 0.13±0.17 <0.005
In-stent area, mm2
 Minimum 7.19±2.62 6.77±2.18 0.42±0.77 <0.05
 Maximum 12.81±4.61 10.48±3.65 2.33±2.25 <0.001
 Mean 9.54±3.48 8.78±2.91 0.82±1.15 <0.01
Percentage of stent malapposed, % 5.5±5.0 19.6±15.1 −14.1±12.4 <0.001
Max stent malapposition, distance in mm 0.57±0.32 1.10±0.34 −0.53±0.33 <0.001
Max stent malapposition, area in mm2 0.88±1.09 2.65±1.88 −1.77±1.35 <0.001

FD-OCT, frequency domain optical coherence tomography; IVUS, intravascular ultrasound.