Skip to main content
. 2015 Dec 22;2(1):e000225. doi: 10.1136/openhrt-2014-000225

Table 4.

Effect of postdilation on stent dimensions assessed by FD-OCT and IVUS (N=6)

IVUS
FD-OCT
Pre Post p Value Pre Post p Value
In-stent diameter, mm
 Minimum 2.55±0.42 2.70±0.50 0.057 2.60±0.41 2.82±0.34 0.036
 Maximum 4.43±0.63 4.77±0.85 0.152 3.84±0.48 4.25±0.65 0.031
In-stent area, mm2
 Minimum 7.73±2.20 7.95±2.72 0.551 7.30±1.62 8.15±1.90 0.023
 Maximum 13.02±3.41 14.77±5.45 0.184 10.72±2.55 13.00±3.62 0.040
Percentage of stent malapposed (%) 7±7 7±4 0.984 34±12 19±10 0.004

FD-OCT, frequency domain optical coherence tomography; IVUS, intravascular ultrasound.