Skip to main content
. Author manuscript; available in PMC: 2017 Feb 1.
Published in final edited form as: Health Phys. 2016 Feb;110(2 0 1):S26–S38. doi: 10.1097/HP.0000000000000462

Table 1.

Relative dose deviation based on the irradiator turntable rotation.

Prescribed dose (Gy) 1.0 2.0 3.0 5.2
ΔD% due to SSD ±20% ±11.7% ±7.4% ±4%
ΔD% due to attenuation ±14% ±7.2% ±5% ±2.8%
Total ΔD% ±24% ±14% ±9% ±5%