Table 3.
Three sets of five needles were inserted into the template at known depths and reconstructed. This comparison gives us the cumulative 3D error from X-ray reconstruction with respect to FTRAC and FTRAC-template registration, i.e. 3D reconstruction error with respect to the template.
Trial number |
Error type |
Error (mm) |
3D error (mm) |
Translation offset (mm) |
||
---|---|---|---|---|---|---|
X | Y | Z | ||||
1 | Absolute | 0.32 | 0.70 | 0.84 | 1.16 | 1.15 |
2 | 0.66 | 0.29 | 1.02 | 1.27 | 1.22 | |
3 | 0.12 | 0.28 | 0.54 | 0.65 | 0.55 | |
1 | Relative | 0.11 | 0.17 | 0.05 | 0.22 | – |
2 | 0.10 | 0.27 | 0.35 | 0.49 | – | |
3 | 0.10 | 0.27 | 0.19 | 0.36 | – |