Skip to main content
. 2015 Dec 30;54(1):153–156. doi: 10.1128/JCM.02440-15

TABLE 2.

Discrepancy testing results

Original illumigene assay resulta and repeat assay result No. of samples
False positive (n = 49)
    Culture positive on repeat testing 2
    Positive by secondary molecular method 34
    Negative by secondary molecular method 12
    Isolate was lost and not further tested 1
False negative (n = 17)
    illumigene assay positive on repeat testing 10
    illumigene assay positive on broth from isolate 6
    Isolate was lost and not further tested 1
a

Original illumigene assay result when culture was used as the gold standard.