TABLE 2.
Discrepancy testing results
Original illumigene assay resulta and repeat assay result | No. of samples |
---|---|
False positive (n = 49) | |
Culture positive on repeat testing | 2 |
Positive by secondary molecular method | 34 |
Negative by secondary molecular method | 12 |
Isolate was lost and not further tested | 1 |
False negative (n = 17) | |
illumigene assay positive on repeat testing | 10 |
illumigene assay positive on broth from isolate | 6 |
Isolate was lost and not further tested | 1 |
Original illumigene assay result when culture was used as the gold standard.