. 2015 Oct 31;408:107–121. doi: 10.1007/s00216-015-9107-2

Table 1.

Limit of detection (LOD) and limit of quantification (LOQ) for real-time PCR (qPCR) and both digital PCR (dPCR) platforms according to the reaction volume analysis. Only the interexperiment variability (repeatability) was taken into account. The LOD was calculated with Probit software. The LOQ was determined as the lowest concentration for which the coefficient of variation was below 25 %. The expected LODs and LOQs are in parentheses

Analysis method Copy number according to analysis system
qPCR QX100 Biomark
LOD LOQ LOD LOQ LOD LOQ
Total reaction volume 2.8 (3) 11–22 5.7 (5.3) 55 (28) 14 (18) 140–210 (100)
Effective reaction size 2.8 (3) 11–22 3.2 (3) 30 (16) 2.4 (3) 14 (18)