TABLE I.

Comparison of thaumatin diffraction obtained from a 1° wedge recorded using an X-ray plate or the crystallography chip.

X-ray plate Crystallography chip
Crystal parameters
Dominant size in sample (μm) 40–50 40–50
Feature width on chip (μm) NA 50
Data collection
Rotation range: total, increment 1 deg, 0.1 deg 1 deg, 0.1 deg
Exposure time (seconds per image) 0.5 0.5
Number of images 10 10
Detector distance (mm) 399.28 271.31
Wavelength (Å) 0.9687 0.9687
Resolution, low (Å) 21.48 16.10
Resolution, high (Å) 1.51 1.55
Number of reflections 1779 2661
Number of reflections, unique 1741 2570
Completeness (%) 4.3 6.8
Multiplicity 1.022 1.035
Rmeas 0.144 0.043
⟨I/σ(I)⟩ 25.0 22.20
Merging space group P 4 2 2 P 4 2 2
Unit cell parameters: a = b, c (Å) 58.39, 151.09 58.37, 151.11