Table 1.

Unloaded and loaded Q values for a 10 × 15 cm surface coil at a variety of operating frequencies. Higher Q ratios correspond to better coil performance. The number of capacitors in the coil was varied with frequency to achieve optimal unloaded Q in each case.

Freq. (MHz) QU QL QU / QL Capacitors
16.3 250 210 1.19 2
32.6 360 215 1.67 2
63.9 350 160 2.19 2
78.6 430 130 3.31 2
123.2 390 75 5.20 4
297.0 115 20 5.75 8