Skip to main content
. Author manuscript; available in PMC: 2016 Jan 15.
Published in final edited form as: Horm Metab Res. 2008 May;40(5):329–337. doi: 10.1055/s-2008-1073156

Table 1.

Sensitivity and specificity of measurements of plasma fractionated metanephrines for diagnosis of pheochromocytoma

Center Method Diagnosis URL (pg/ml) Diag Accuracy Ref.
P-Yes P-No NMN MN Sens Spec
NIH HPLC 214 644 112 61 99 % 89 % [8]
Vienna HPLC 17 14 121 61 100 % 100 % [7]
Mayo Clinic HPLC 56 445 165 99 96 % 85 % [9]
University of Essen RIA 24 126 126 37 96 % 79 % [10]
Sternberk Hospital HPLC 25 1 235 112 61 100 % 97 % [11]
All Centers 336 2 464 98 % 92 %

Abbreviations: P-Yes: number of patients with pheochromocytoma; P-No: number of patients in whom pheochromocytoma was excluded; URL: upper reference limits of reference intervals; NMN: normetanephrine; MN: metanephrine; Diag Accuracy: Diagnostic accuracy; Sens: Diagnostic sensitivity; Spec: Diagnostic specificity