Table 3.

Drilling accuracy data for medial drilling experiments

Parametersa,b Targeting error magnitude (mm)
X-Y position of error (mm)
μ ± σ RMS Max. X, μ ± σ Y, μ ± σ
0°/6 mm 0.035 ± 0.011 0.037 0.060 −0.031 ± 0.014 0.012 ± 0.007
15°/6 mm 0.027 ± 0.009 0.029 0.042 −0.019 ± 0.011 0.094 ± 0.017
30°/6 mm 0.028 ± 0.021 0.035 0.071 −0.023 ± 0.024 0.003 ± 0.012
45°/6 mm 0.053 ± 0.023 0.058 0.090 −0.028 ± 0.021 −0.041 ± 0.020
60°/6 mm 0.098 ± 0.036 0.105 0.166 −0.056 ± 0.029 −0.079 ± 0.028
75°/6 mm 0.116 ± 0.011 0.117 0.131 −0.037 ± 0.021 −0.109 ± 0.012
45°/3 mm 0.072 ± 0.017 0.074 0.104 −0.052 ± 0.022 −0.046 ± 0.011
45°/9 mm 0.117 ± 0.043 0.125 0.185 −0.054 ± 0.022 −0.103 ± 0.041
a

12 trials were performed for each case

b

Parameters are angle of bone at contact point/cantilevered length of drill bit from end of bushing to contact point