Table 2.
accuracy (n = 3) | relative standard deviation (n = 30) |
|||||||
---|---|---|---|---|---|---|---|---|
analyte | calibration standard range (µg) | low | medium | high | smoke matrix effecta (n = 3) | LOD (µg) (n = 7) | 3R4F ISOb | 3R4F CIc |
formaldehyde | 5–300 | 106% | 102% | 99% | 0% | 2.4 | 19% | 19% |
acetaldehyde | 50–3000 | 90% | 100% | 98% | −1.3% | 2.7 | 10% | 12% |
acetone | 25–1500 | 102% | 99% | 95% | 3.1% | 6.4 | 14% | 14% |
acrolein | 5–300 | 100% | 103% | 95% | −7.5% | 0.1 | 10% | 13% |
propionaldehyde | 5–300 | 83% | 102% | 103% | −2.4% | 0.6 | 13% | 16% |
crotonaldehyde | 2.5–150 | 101% | 101% | 96% | −3.6% | 0.2 | 12% | 13% |
methyl ethyl ketone | 5–300 | 104% | 93% | 96% | −7.8% | 0.3 | 14% | 13% |
Matrix effects were assessed by comparing slopes of calibration curves prepared from two set of standards. One set was prepared neat, and the second set was prepared with smoke matrices present. “+” or “−” indicates a matrix-enhanced or -suppressed effect, respectively.
ISO (35 mL puff volume, 60 s puff interval, 2 s puff duration, and filter-tip vent open).
CI (55 mL puff volume, 30 s puff interval, 2 s puff duration, and 100% vent block).