Published in final edited form as: J Magn Reson Imaging. 2015 Jul 27;43(2):414–425. doi: 10.1002/jmri.24999

Table 2.

Measured T1 values by BLESSPC fitting in a set of FLASH-MOLLI phantom experiments (group I) with T1s 236 – 1852 ms and emulated HRs 40 – 100 bpm.

Phantom index Ref T1(ms) Ref T2(ms) Measured T1 (ms) by BLESSPC fitting
HR=40 HR=60 HR=80 HR=100 Mean
1 236 45 234 232 233 233 233±0.6
2 455 40 450 449 449 449 449±0.6
3 575 40 572 572 572 572 572±0.1
4 658 67 668 667 666 665 667±1.5
5 702 36 692 692 691 691 692±0.4
6 1208 71 1223 1220 1220 1220 1221±1.4
7 1459 39 1422 1425 1427 1427 1425±2.5
8 1553 63 1560 1559 1560 1561 1560±0.8
9 1597 38 1560 1562 1562 1566 1563±2.6
10 1602 67 1601 1602 1600 1600 1601±0.8
11 1852 67 1857 1853 1852 1852 1854±2.4