Published in final edited form as: Magn Reson Med. 2015 Jul 21;76(1):20–31. doi: 10.1002/mrm.25828

Table 1.

Parallel excitation RF pulse power comparison with different constraints. For conciseness, values are listed for the channel with the highest power level for the given measurement (indicated in italics). Good agreement between the fully constrained power predictions and measurements can be seen in columns three and four.

Measured Predicted


Unconstrained Global SAR
Constrained*
Fully
Constrained**
Fully
Constrained
Global SAR (W/kg) 5.1 2.9 2.3 2.4
FWD Peak (W) 1103.2 (ch3) 916.2 (ch1) 683.9 (ch4) 700.0
RFL Peak (W) 86.5 (ch3) 67.5 (ch3) 46.4 (ch3) 50.0
FWD Average (W) 65.3 (ch1) 40.3 (ch1) 30.7 (ch1) 33.5
RFL Average (W) 3.7 (ch2) 2.2 (ch3) 1.6 (ch3) 2.0

The following constraints were implemented:

*

global SAR ≤ 3.2 W / kg

**

global SAR ≤ 3.2 W / kg, Pfwd,peakl700W,Prfl,peakl50W,Pfwd,avel50W,Prfl,avel25W