Table 2.

Maximum averaged 10 g SAR and excitation fidelity results of RF pulses designed with different constraints.

Pulse
Design
Type
Unconstrained Global SAR
Constrained
Fully
Constrained
Fully
Constrained,
Averaged
10g SAR:
0.5W/kg
Averaged
10g SAR:
0.5W/kg
Fully
Constrained,
Averaged 10g
SAR:
0.025W/kg
Maximum Average 10g SAR (W/kg) 1.18 0.82 0.64 0.45 0.45 0.02
Predicted Maximum Peak Power (W) 1088 992 700 700 754 75
NRMSE 0.022 0.023 0.026 0.032 0.032 0.1

Lower maximum average 10g SAR results on the designed RF pulses are obtained compared to enforced limits due to the predefined maximum worst-case 10g SAR overestimations during VOP calculations.