Skip to main content
. 2016 Jan 21;10:1. doi: 10.1186/s13065-016-0148-1

Table 2.

Sensor performance at different time points on the CA3-CA1 response

Time/s 0.000–0.025 0.025–0.050 0.050–0.075 0.075–0.100 0.100–0.125 0.125–0.150 0.150–0.175 0.175–0.200 0.200–0.225 0.225–0.250
False positives 1 2 2 1 1 1 1 1 1 0
True negatives 172 168 169 168 166 165 165 167 165 165
True positives 14 95 95 68 39 30 22 20 15 14
False negatives 328 248 242 264 299 300 312 313 313 309
% Sensitivity 4 28 28 20 12 9 7 6 5 4
% Specificity 99 99 99 99 99 99 99 99 99 100
% Accuracy 36 51 52 47 41 39 37 37 36 37

Note that outlier responses were removed from the analysis. Outliers were defined as (1) did not trigger; (2) excessively noisy response resulting in atypical transient shape and (3) statistical outlier for each set of 12 sensors tested per sample, defined as outside 1.5x the interquartile range from the median current response