Precision, accuracy, and recovery for DA.

Norminal Concentration (μg/L) Intra-Day Inter-Day Analyte Recovery (%, n = 6) Matrix Effect (%, n = 3)
RSDr (%) Biases (%) RSDR (%) Biases (%)
0.37 7.6 −6.5 10.4 −7.4 88.4 ± 8.9 96.8 ± 2.1
0.74 5.8 −4.7 12.7 −5.3 92.8 ± 5.3 97.7 ± 2.6
3.7 2.1 −1.9 3.6 −3.1 97.2 ± 2.7 100.9 ± 1.4
37 2.4 −3.1 2.6 −4.2 102.5 ± 4.3 101.7 ± 1.2