CT abdominal imaging protocols

Patient Scan protocol
Paediatric abdomen (kg)  
 0–20 Helical scan, 80 kVp, 50 mAs, detector collimation 64 × 0.625 mm, rotation time 0.5 s, slice thickness 3 mm, pitch 0.891 and iDose level 4 iterative reconstruction
 20–40 Helical scan, 80 kVp, DoseRight AEC, rotation time 0.5 s, detector collimation 64 × 0.625 mm, slice thickness 3 mm, pitch 0.891 and iDose level 4 iterative reconstruction
 40–60 Helical scan, 100 kVp, DoseRight AEC, rotation time 0.5 s, detector collimation 64 × 0.625 mm, slice thickness 3 mm, pitch 0.891 and iDose level 4 iterative reconstruction
 >60 Helical scan, 120 kVp, DoseRight AEC, rotation time 0.5 s, detector collimation 64 × 0.625 mm, slice thickness 3 mm, pitch 0.891 and iDose level 4 iterative reconstruction

AEC, automatic exposure control.