Solution type | Acceptance criteria and comments |
---|---|
Calibration blank | To check the background contamination. There should not be any contamination |
Sensitivity check standard | For determination of LOD and LOQ of the instrument |
Standards (CRMs) | To check the linearity range of the instrument |
Calibration blank | No contamination should be there in blank |
Calibration verification standard | It should read between 80 and 120 % |
Reagent blank | Should be lower than 10 times the acceptable limit of error for that particular analyte |
Standard reference material | By matrix spiking. Recovery between 80 and 120 % |
Establishment of LOD/LOQ for the method used | By replicate analysis of spiked samples at ppb level and after getting the LOQ value, practically demonstrated to get confidence in the method used |
Samples | Absorbance should be <highest standard % RSD of two readings should be ±20 % Recovery of spiked sample should be 80–120 % Values less than LOQ evaluated |
SRM and calibration blank | Used for checking the instrumental performance in between the sample analysis |