Solution type Acceptance criteria and comments
Calibration blank To check the background contamination. There should not be any contamination
Sensitivity check standard For determination of LOD and LOQ of the instrument
Standards (CRMs) To check the linearity range of the instrument
Calibration blank No contamination should be there in blank
Calibration verification standard It should read between 80 and 120 %
Reagent blank Should be lower than 10 times the acceptable limit of error for that particular analyte
Standard reference material By matrix spiking. Recovery between 80 and 120 %
Establishment of LOD/LOQ for the method used By replicate analysis of spiked samples at ppb level and after getting the LOQ value, practically demonstrated to get confidence in the method used
Samples Absorbance should be <highest standard
 % RSD of two readings should be ±20 %
Recovery of spiked sample should be 80–120 %
Values less than LOQ evaluated
SRM and calibration blank Used for checking the instrumental performance in between the sample analysis