TABLE 4.

Agreement between the GeneXpert Ebola assay and the L gene qRT-PCR on the LightCycler Nano and the SmartCycler at different CT value ranges, using clinical samples collected from suspected EVD cases in Freetown, Sierra Leone from 2014 to 2015

SmartCycler L gene qRT-PCR CT value range (no. of samples) GeneXpertEbola assaya
LightCycler Nano L gene qRT-PCRa
GP target
NP target
One or both targets (GP and NP)
+ + + +
CT < 30 (n = 81) 81 0 81 0 81 0 81 0
100%b 100% 100% 100%
CT 30–34.99 (n = 15) 15 0 15 0 15 0 15 0
100% 100% 100% 100%
CT 35–39.99 (n = 16) 9 7 15 1 15 1 15 1
56.25% 93.75% 93.75% 93.75%
CT 40–45 (n = 10) 2 8 9 1 10 0 9 1
20.0% 90.0% 100% 90.0%
No fluorescence signal above the threshold (n = 159) 2 157 3 156 4 155 1 158
98.74% 98.11% 97.48% 99.37%
a

+, Positive; −, negative.

b

Percentage agreement between applicable assay and SmartCycler L gene qRT-PCR using a cutoff CT of 45.