. 2016 Jan 28;54(2):412–422. doi: 10.1128/JCM.02469-15

TABLE 1.

Test accuracy of MAC-ELISA using VLP and/or NS1 antigensa

Control panel by disease panel Antigen ELISA result True status
AUC (95% CI) % sensitivity (95% CI) % specificity (95% CI)
Disease No disease
JEV
    Other groupsb JEV VLP Positive 16 5 1.00 (1.00–1.00) 100.00 (79.41–100.00) 92.54 (83.44–97.53)
Negative 0 62
JEV NS1 Positive 16 0 1.00 (1.00–1.00) 100.00 (79.41–100.00) 100.00 (94.64–100.00)
Negative 0 67
JEV VLP and NS1 Positive 16 0 NAc 100.00 (79.41–100.00) 100.00 (94.64–100.00)
Negative 0 67
    WNV and other groups JEV VLP Positive 16 23 0.99 (0.99–1.00) 100.00 (79.41–100.00) 82.44 (74.83–88.53)
Negative 0 108
JEV NS1 Positive 16 41 0.99 (0.96–1.01) 100.00 (79.41–100.00) 68.70 (60.02–76.52)
Negative 0 90
JEV VLP and NS1 Positive 16 13 NA 100.00 (79.41–100.00) 90.08 (83.63–94.61)
Negative 0 118
WNV
    Other groups WNV VLP Positive 64 0 1.00 (1.00–1.00) 100.00 (94.40–100.00) 100.00 (94.64–100.00)
Negative 0 67
WNV NS1 Positive 64 2 0.99 (0.98–1.01) 100.00 (94.40–100.00) 97.01 (89.63–99.63)
Negative 0 65
WNV VLP and NS1 Positive 64 0 NA 100.00 (94.40–100.00) 100.00 (94.64–100.00)
Negative 0 67
    JEV and other groups WNV VLP Positive 64 8 0.99 (0.98–1.00)d 100.00 (94.40–100.00) 90.36 (81.89–95.75)
Negative 0 75
WNV NS1 Positive 64 16 0.89 (0.83–0.94) 100.00 (94.40–100.00) 80.72 (70.59–88.56)
Negative 0 67
WNV VLP and NS1 Positive 64 8 NA 100.00 (94.40–100.00) 90.36 (81.89–95.75)
Negative 0 75
a

Accuracy was defined as the ability to distinguish the disease panel (JEV or WNV serum panels) from the control panel using the positive-cutoff criterion (P/N ≥ 3.0) as the evidence of infection.

b

Other groups include DENV, YFV, Zika, HTN, CHIKV, and negative serum panels.

c

NA, not available.

d

P < 0.05.