Table 2.
Basic Setting Parameters in the Dispensing Programs and Spotting
| Tip setting | |
|---|---|
| Aspiration volume (µl) | 3 |
| Aspiration flow rate (µl/sec) | 0.20 |
| Return sample in source plate after spotting | No |
| Return sample in source plate if failed stroboscope check | No |
| Delay time (sec) | 1.0 |
| Before aspiration dry pad time (sec) | 0.10 |
| after aspiration dry pad time (sec) | 0.0 |
| Piezo voltage (V) | 100 |
| Piezo pulse width (µsec) | 50 |
| Piezo frequency (100 Hz) | 100 |
| Wash conditions | |
| Wash time | 30.00 |
| Dry time (sec) | 0.10 |
| Extrawash well | Extrawash |
| Extrawash volume | 0.00 |
| Extrawash flow rate (µl/sec) | 0.50 |
| Extrawash piezo | No |
| Stroboscope details | |
| Delay before check | 0.50 |
| Checks | 4 |
| Delay between checks (sec) | 0.10 |
| % CV tolerance across checks | 5% |
| Z heights | |
| Z height from target | 0.3 |
| Z height from traverse target | 0.0 |