Table 5.

Stability and dilution integrity data of PTB.

Stability Nominal conc. (ngmL−1) Plasma Urine
Measured conc. (ngmL−1  ±  SD) Precision (CV %) Accuracy (R  %) Measured conc. (ngmL−1  ±  SD) Precision (CV %) Accuracy (R  %)
Bench top (6 h) 50 46.01 ± 0.42 0.913 92.01 ± 0.84 47.67 ± 0.40 0.848 95.35 ± 0.81
100 92.08 ± 0.68 0.736 92.08 ± 0.68 95.63 ± 1.53 1.605 95.63 ± 1.53

Freeze thaw (3 cycles) 50 50.67 ± 0.31 0.606 101.34 ± 0.61 50.00 ± 1.00 2.016 100.0 ± 2.02
100 100.47 ± 1.08 1.071 100.47 ± 1.08 100.0 ± 1.07 1.074 100.0 ± 1.0

14 days at −80°C 50 42.94 ± 0.91 2.108 85.87 ± 0.91 47.20 ± 0.953 2.019 94.40 ± 1.91
100 86.67 ± 0.68 0.788 86.67 ± 0.68 94.11 ± 1.26 1.343 94.11 ± 1.26

Dilution integrity 45 43.54 ± 0.61 1.393 96.76 ± 1.35 44.48 ± 0.59 1.337 98.84 ± 1.32
90 90.21 ± 1.66 1.835 100.23 ± 1.84 88.27 ± 0.73 0.824 98.08 ± 0.81