. 2015 Sep 30;10(1):240–241. doi: 10.1177/1932296815608871

Table 1.

System Accuracy Results for GlucoCheck neo and Element.

Results within minimum accuracy criteria of
ISO 15197:2013b
System Lot Comparison method Study date ISO 15197:2003a Criterion A Criterion B
GluNeo, tested as GlucoCheck neo 1 Cobas c111 April-May 2013 198/200 (99%) 193/200 (96.5%) 600/600 (100%)
2 195/200 (97.5%) 190/200 (95%)
3 199/200 (99.5%) 195/200 (97.5%)
Element 1 Cobas Integra 400 Plus Oct 2014 198/200 (99%) 195/200 (97.5%) 600/600 (100%)
2 198/200 (99%) 193/200 (96.5%)
3 196/200 (98%) 191/200 (95.5%)
a

At least 95% of results of each individual test strip lot have to be within ±15 mg/dl of comparison measurement results at glucose concentrations <75 mg/dl and within ±20% at glucose concentrations ≥75 mg/dl.

b

For criterion A, at least 95% of results of each individual test strip lot have to be within ±15 mg/dl of comparison measurement results at glucose concentrations <100 mg/dl and within ±15% at glucose concentrations ≥100 mg/dl. For criterion B, at least 99% of results of all 3 test strip lots combined have to be within consensus error grid zones A and B.