Table 1. Absolute maximum temperature increase measured between the electrodes and at the tip of the electrodes.
Median maximum increase in temperature in°C (range) in tissue phantom | |||
---|---|---|---|
No stent | Stent parallel | Stent perpendicular | |
Between electrodes | |||
90 pulses | 10.2 (7.4–12.6) | 8.6 (6.9–11.5)* | 10.7 (9.6–11.7)* |
270 pulses | 23.6 (17.0–26.7) | 21.0 (19.7–23.5)** | 22.0 (21.1–24.0)** |
Electrode tip | |||
90 pulses | 13.3 (11.6–14.1) | 19.4 (14.7–21.3)* | 23.2 (19.8–25.2)* |
270 pulses | 24.8 (17.2–26.6) | 26.5 (25.5–35.4)** | 33.1 (33.1–38.1)** |
Data represent the median and range of 5 experiments.
* 4 experiments, 1 aborted due to high current,
** 3 experiments, 2 aborted due to high current.