Table 3.

Plan comparison for planning target volume of gross tumour volume (PGTV), planning target volume (PTV)1, PTV2, monitor units (MUs) and delivery time

Parameter Intensity-modulated radiation therapy, mean ± SD (Gy) SmartArc-based volumetric modulated arc therapy, mean ± SD (Gy) p-value
PGTV
D98% 65.2 ± 2.9 64.5 ± 3.4 0.140
D2% 73.5 ± 1.2 72.2 ± 1.2 0.001
Dmean 70.7 ± 0.9 70.2 ± 0.6 0.045
 CI 0.54 ± 0.12 0.59 ± 0.12 <0.001
 HI 1.09 ± 0.03 1.05 ± 0.02 <0.001
PTV1
D98% 58.7 ± 1.9 58.8 ± 2.3 0.898
D2% 69.2 ± 2.1 68.5 ± 1.9 0.036
Dmean 65.2 ± 1.1 64.7 ± 0.8 0.023
 CI 0.42 ± 0.07 0.49 ± 0.08 <0.001
 HI 1.17 ± 0.02 1.14 ± 0.03 <0.001
PTV2
D98% 51.9 ± 3.2 52.3 ± 2.9 0.510
D2% 65.2 ± 3.8 64.4 ± 3.8 0.059
Dmean 58.9 ± 1.3 58.4 ± 1.2 0.009
 CI 0.47 ± 0.05 0.51 ± 0.07 <0.001
 HI 1.18 ± 0.02 1.16 ± 0.03 0.001
MUs 741.9 ± 48.2 625.7 ± 66.6 <0.001
Time (min) 14.2 ± 1.1 6.3 ± 0.4 <0.001

CI, conformity index; D2%, near maximum dose; D98%, near minimum dose; Dmean, mean dose; HI, homogeneity index; SD, standard deviation.