Table 2.
Operating conditions of the ICP-MS and the chromatographic systems.
Inductively coupled plasma mass spectrometry (ICP-MS) | |
---|---|
Radio frequency power | 1300 W |
Sampler and skimmer cones | Nickel |
Argon flow rates: | |
Plasma | 15 L/min |
Auxiliary | 0.86 L/min |
Nebulizer | 0.96 L/min |
Date acquisition mode | Graphics (signal intensity versus time) |
Analytical mass (amu) | 75As |
Chromatography | |
Anion exchange | |
Guard column | IonPac AG 19 (50 × 4 mm) |
Analytical column | IonPac AS 19 (250 × 4 mm) |
Mobile phase | 10 mmol/L CH3COONa, 3 mmol/L KNO3, 2 mmol/L NaH2PO4, and 0.2 mmol/L EDTA-2Na adjusted to pH 10.7 with 4% NaOH |
Flow rate | 1.0 mL min−1 |
Injection volume | 30 μL |