Skip to main content
. Author manuscript; available in PMC: 2017 Jul 1.
Published in final edited form as: Nicotine Tob Res. 2015 Aug 9;18(7):1575–1580. doi: 10.1093/ntr/ntv162

Table 2.

Method validation for menthol including precision and accuracy.

Spike Level Average (μg/g) Std. Dev. (μg/g) Calculated Conc. (μg/g) Precision (CV) Accuracy (Recovery)
1 μg/g 1.00 0.13 1.01 13% 99.0%
5 μg/g 5.02 0.21 5.05 4.2% 99.5%
252 μg/g 277.5 5.60 252.3 2.02% 110.0%
757 μg/g 856.4 14.7 756.8 1.72% 113.2%
5045 μg/g 5479.1 128.5 5045.4 2.35% 108.6%