TABLE 2.

Effectiveness of Various Prion Decontamination Methods Evaluated by Selected Inactivation Studies

Publication Year Prion
strain
Source
material
Assay
used
Decontamination
methods
Transmission,
% (infected/total)
Log
reduction
Taylor et al10 1997 22A scrapie strain Brain macerates VM mice, IC Immersed in H2O, 1 h then autoclaving in H2O, 121°C; 30 min 100 (11/11) ...
Autoclaving in NaOH (2M), 121°C; 30 min 0 (0/18) ...
Taylor et al11 1998 263K scrapie strain Brain macerates Hamsters, homogenate, IC Autoclaving, 134°C; 18 min (1 cycle) 100 (10/10)a 3.3
Autoclaving, 134°C; 18 min (2 cycles) 60 (3/5)a 4.6
Fichet et al14 2004 263K scrapie strain Stainless steel wires in 10% homogenate Hamsters, implanted NaOCl (20,000 ppm), 20°C; 1 h 0 (0/8) >5.6
NaOH (1N), 20°C; 1 h 0 (0/12) >5.6
Autoclaving (dry),b 134°C; 18 min 60 (6/10) 4 to 4.5
Autoclaving in H2O, 134°C; 18 min 0 (0/11) > 5.6
Yan et al15 2004 263K scrapie strain Stainless steel wires in 10% homogenate Hamsters, implanted Autoclaving, 134°C; 18 min 10 (1/10) ...
NaOH (1 M) bath, 24 h then autoclaving 134°C; 18 min 20 (2/10) ...
Enzymatic detergent (2%) wash then autoclaving, 134°C; 18 min 100 (10/10) ...
Alkaline detergent (pH, 11) wash 70°C; 10 min then autoclaving, 134°C; 18 min 22 (2/9) ...
Jackson et al16 2005 RML strain Stainless steel wires in 10% or 20% homogenate Tg20 mice, implanted Autoclaving, 121°C; 20 min (20% homogenate) 33 (2/6)c ...
Autoclaving, 134°C; 20 min (20% homogenate) 25 (1/4)c ...
Autoclaving, 134°C; 20 min (10% homogenate) 100 (13/13) ...
Peretz et al22 2006 Sporadic CJD strain Stainless steel wires in 10% homogenate Tg23372 mice, implanted Autoclaving, 121°C; 30 min 0 ...
Autoclaving, 121°C; 2 h 73 ...
Autoclaving, 134°C; 2 h 46 ...
Fernie et al17 2007 301V BSE, 263K scrapie, or 22A scrapie strains Brain macerates VM mice or LVG hamsters, homogenate, IC Autoclaving, 134°C; 18 min (301V, VM mice) 64 (9/14) ...
Autoclaving, 134°C; 18 min (263K, LVG hamsters) 13 (2/16) ...
Autoclaving, 134°C; 18 min (22A, VM mice) 0 (0/15) ...
Giles et al21 2008 301V BSE, cattle BSE Stainless steel wires in 10% homogenate Tg2091 or Tg4092 mice, implanted Autoclaving, 134°C; 2 h (301V strain, Tg 2091 mice) 14 ...
Autoclaving, 134°C; 2 h (cattle BSE strain, Tg 4092 mice) 89 ...
Lehmann et al18 2009 263K scrapie strain Stainless steel wires in 20% homogenate Hamsters, implanted Autoclaving, 134°C; 18 min 57 4.11
Rogez-Kreuz et al19 2009 263K scrapie strain Stainless steel wires in 10% homogenate Hamsters, implanted Autoclaving, 134°C; 18 min 50 ≥5 to 6
NaOH (1 N), room temperature, 1 h then autoclaving, 134°C: 18 min 28 ≥5 to 6
Enzymatic detergent (2%), 37°C; 10 min then autoclaving, 134°C; 18 min 100 4.0
Alkaline detergent A (1%), 70°C; 10 min then autoclaving, 134°C; 18 min 0 ≥5 to 6

NOTE. BSE, bovine spongiform enchephalopathy; h, hours; IC, intracerebrally inoculated; LVG, Lakeview Golden; min, minutes.

a

Transmission rates are from hamsters injected with 10−1 dilution group.

b

Wires were placed on support during autoclaving at 134°C.

c

All infected mice did not display clinical signs and survived to the end of the experiment but were classified as infected after neuropathological examination.