. 2016 Feb 12;10(2):e0004443. doi: 10.1371/journal.pntd.0004443

Table 3. Analytical sensitivity of HtLAMP-Pv, compared with other published P. vivax LAMP primers, using a DNA dilution series of clinical sample with qPCR confirmed parasitemia.

Each sample was tested in duplicate in the HtLAMP platform using each of the three P. vivax LAMP primer sets. The dilution at which both duplicates were positive was the limit of detection for each of the primer sets in the HtLAMP platform (Pos* indicates dilutions at which only one of the duplicates was positive). HtLAMP-Pv was able to detect 1.4 parasites/ μL compared with previously published P. vivax LAMP primers.

Sample Quantitative PCR (parasites/μL calculated) [23] 18s rRNA target [30] Pvr64 target HtLAMP-Pv cox1 target
1 Original 90,000 Pos Pos Pos
2 45,000 Pos Pos Pos
3 22,500 Pos Pos Pos
4 11,250 Pos Pos Pos
5 5625 Pos Pos Pos
6 2813 Pos Pos Pos
7 1406 Pos Pos Pos
8 703 Pos Pos Pos
9 352 Pos* Pos Pos
10 176 Pos Pos
11 88 Pos
12 44 Pos
13 22 Pos
14 11 Pos
15 5.5 Pos
16 2.7 Pos* Pos
17 1.4 Pos
18 0.7 Pos*
19 0.3