Skip to main content
. 2015 Dec;56(4):E190–E195.

Tab. I.

Analytical data of the acrylamide standard solution analysed by HPLC-UV: linear regression equation (peak area/conc.) and correlation coefficient (R2) calculated in the linearity range 0.05-2 μg/g at wavelengths 210 and 200.5 nm; limit of detection (LOD, s/n = 3/1) expressed as μg/kg; limit of quantification (LOQ, s/n = 7/1) expressed as μg/kg; intra-day RSD and inter-day RSD (n = 3).

Linear regression equation R2 LOD (μg/kg) LOQ (μg/kg) Intra -day a RSD Inter-day b RSD Wavelength
y = 116616 x -4415. 8 0.998 10 28 3-6.6 6.8 210 nm
y = 250771 x -2640. 9 0.999 10 28 4.2-7.2 1.9 200.5 nm
a

Values are referred to six independent samples analyzed in one day

b

Values are referred to six independent samples analyzed three times in three different days