Skip to main content
Journal of Clinical Pathology logoLink to Journal of Clinical Pathology
. 1975 Aug;28(8):625–630. doi: 10.1136/jcp.28.8.625

Measurement of plasma renin activity by semi-micro radioimmunoassay of generated angiotensin I.

M J Dillon
PMCID: PMC475790  PMID: 1184761

Abstract

A semi-micro method for determination of plasma renin activity (PRA) by radioimmunoassay of generated angiotensin I (AI) is described. The method permits measurement of PRA on 250 mul of plasma without loss of specificity, sensitivity, accuracy or preision. The small sample size has considerable application in terms of investigation of infants and young children. A reference range was established for healthy children on free diets. There was a 15-fold decline in PRA with age from a mean value of 1404pgAI/ml h-1 (ngAI/1. h-1) in infancy to a mean of 85 pgAI/ml h-1 (ngAI/1. h-1) in adult life.

Full text

PDF

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Boyd G. W., Fitz A. E., Adamson A. R., Peart W. S. Radioimmunoassay determination of plasma-renin activity. Lancet. 1969 Feb 1;1(7588):213–218. doi: 10.1016/s0140-6736(69)91237-9. [DOI] [PubMed] [Google Scholar]
  2. Brech W., Franz H. E. Die radioimmunologische Bestimmung von Angiotensin I zur Messung der Reninaktivität. Klin Wochenschr. 1972 Mar 15;50(6):327–331. doi: 10.1007/BF01485180. [DOI] [PubMed] [Google Scholar]
  3. Cohen E. L., Grim C. E., Conn J. W., Blough W. M., Jr, Guyer R. B., Kem D. C., Lucas C. P. Accurate and rapid measurement of plasma renin activity by radioimmunoassay. Results in normal and hypertensive people. J Lab Clin Med. 1971 Jun;77(6):1025–1038. [PubMed] [Google Scholar]
  4. Dillon M. J., Ryness J. Proceedings: Plasma renin activity and aldosterone concentration in children. Arch Dis Child. 1974 Oct;49(10):823–823. doi: 10.1136/adc.49.10.823-a. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Haber E., Koerner T., Page L. B., Kliman B., Purnode A. Application of a radioimmunoassay for angiotensin I to the physiologic measurements of plasma renin activity in normal human subjects. J Clin Endocrinol Metab. 1969 Oct;29(10):1349–1355. doi: 10.1210/jcem-29-10-1349. [DOI] [PubMed] [Google Scholar]
  6. Hollemans H. J., van der Meer J., Kloosterziel W. Measurement of renin activity in plasma by a radioimmunoassay for angiotensin I. Clin Chim Acta. 1969 Jun;24(3):353–357. doi: 10.1016/0009-8981(69)90105-3. [DOI] [PubMed] [Google Scholar]
  7. Katz F. H., Smith J. A. Radioimmunoassay of angiotensin I: comparison of two renin activity methods and use for other measurements of the renin system. Clin Chem. 1972 Jun;18(6):528–533. [PubMed] [Google Scholar]
  8. Kodish M. E., Katz F. H. Plasma renin concentration: comparison of angiotensinase inhibitors and correlation with plasma renin activity and aldosterone. J Lab Clin Med. 1974 May;83(5):705–715. [PubMed] [Google Scholar]
  9. Kurtz A. B. Plasma-renin activity determined by a labelled antibody assay. Horm Metab Res. 1971 May;3(3):203–206. doi: 10.1055/s-0028-1094157. [DOI] [PubMed] [Google Scholar]
  10. Malvano R., Zucchelli G. C., Rosa U., Salvetti A. Measurement of plasma renin activity by angiotensin I radioimmunoassay: (I) an assessment of some methodological aspects. J Nucl Biol Med. 1972 Jan-Mar;16(1):24–31. [PubMed] [Google Scholar]
  11. Menard J., Catt K. J. Measurement of renin activity, concentration and substrate in rat plasma by radioimmunoassay of angiotensin I. Endocrinology. 1972 Feb;90(2):422–430. doi: 10.1210/endo-90-2-422. [DOI] [PubMed] [Google Scholar]
  12. Page L. B., Dessaulles E., Lagg S., Haber E. Interference with immunoassays of angiotensin I and II by proteins in human plasma. Clin Chim Acta. 1971 Aug;34(1):55–62. doi: 10.1016/0009-8981(71)90065-9. [DOI] [PubMed] [Google Scholar]
  13. Rössler R., Hornef W., Klaus D., Simsch A. Bestimmung der Plasmareninaktivität durch radioimmunologischen Nachweis von Angiotensin 1. Vergleich mit der biologischen Methode. Klin Wochenschr. 1971 Aug 1;49(15):870–876. doi: 10.1007/BF01513470. [DOI] [PubMed] [Google Scholar]
  14. Sealey J. E., Gerten-Banes J., Laragh J. H. The renin system: Variations in man measured by radioimmunoassay or bioassay. Kidney Int. 1972 Apr;1(4):240–253. doi: 10.1038/ki.1972.34. [DOI] [PubMed] [Google Scholar]
  15. Sealey J. E., Laragh J. H. Searching out low renin patients: limitations of some commonly used methods. Am J Med. 1973 Sep;55(3):303–314. doi: 10.1016/0002-9343(73)90132-0. [DOI] [PubMed] [Google Scholar]
  16. Viol G. W., Keane P. M., Speed J. F., Smith E. K. Assay of plasma renin activity using commercially available reagents. Clin Biochem. 1972 Dec;5(4):251–265. doi: 10.1016/s0009-9120(72)80041-9. [DOI] [PubMed] [Google Scholar]
  17. Zucchelli G. C., Malvano R., Rosa U., Salvetti A. Control of the enzyme systems in plasma renin activity measurement by angiotensin I radioimmunoassay. J Nucl Biol Med. 1973 Oct-Dec;17(4):187–195. [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES