Table B1.

Results for three example scintillators described in table 2 and photodetector timing jitter 0.2 ns fwhm. See text and table A1 for variable definitions.

LSO LSO LSOa LaBr3 : Ce LaBr3 : Ce Ultra-
fast
Ultra-
fast
Npe (total) 1000 4000 4000 1900 7600 1000 4000
Average Npe (photodetector A) 575 2302 2302 1030 4120 575 2301
Average P (photodetector A) 36.9 140.5 357 115.8 452.8 364.9 1457
Average Npe (photodetector B) 424 1697 1697 869 3479 424 1698
Average P (photodetector B) 27.7 104.3 263 98.2 383.0 268.8 1075
Fopt 0.018 0.018 0.0024 0.022 0.022 0.065 0.065
P Fopt (photodetector A) 0.66 2.5 0.86 2.5 10.0 23.7 95
P Fopt (photodetector B) 0.50 1.9 0.63 2.3 8.4 17.4 70
Average HA (photodetector A) 2.1 6.9 8.3 6.1 22 59 235
Average HB (photodetector B) 1.7 5.1 6.0 5.2 19 43 169
WWAB (ns fwhm) 0.242 0.111 0.109 0.146 0.072 0.041 0.020
a

SER rise time 1 ns and decay time 10 ns used in figure B1(b). All other columns had SER rise time 0.2 ns and decay time 2 ns.